How to charge
13
LED and state Explanation
Status LEDs
shining red
Wrong polarity of corresponding hearing
instrument.
Check if battery is inserted correctly.
Check if hearing instruments are
inserted correctly in charging slots.
Status LEDs
blinking red
Wrong battery in left/right hearing
instrument. Charging can not be
initiated.
Check type of batteries.
Status LEDs
blinking orange
External temperature is too high. The
charging process is interrupted to protect
the functionality of the rechargeable
batteries. Charging continues when the
external temperature drops under 45°C
(113°F).
Place the eCharger in a cooler
location.
Consult your Hearing Care Professional if you encounter
problems.